| Electrical Requirements: | 120 VAC ± 10% (132 VAC/108 VAC), 60 Hz |
| Rated Current Input: | Platinum 5 and XL: 4.3 A
Platinum 10: 4.7 A |
| Sound Level: | Platinum 5 and XL: 50 dB Average
Platinum 10: 58 dB Average |
| Altitude: | Platinum 5 and XL - Up to 6,000 ft (1828 meters) above sea level without degradation of concentration levels. From 6,000 ft (1828 meters) to 13,129 ft (4000 meters) below 90% efficiency
Platinum 10 - Up to 4,000 ft (1230 meters) above sea level without degradation of concentration levels. From 4,000 ft (1230 meters) to 13,129 ft (4000 meters) below 90% efficiency |
| *Oxygen Output Concentration Levels: | All 5LXO2/5LX models 87% to 95.6% for flows ranging 0.5 to 5 L/min and all 10LXO2/10LX models 87% to 95.6% for flows ranging 2 L to 10 L/min |
| Maximum Outlet Pressure: | Platinum 5 and XL: 5 psi ± 0.5 psi (34.5 kPa ± 3.45 kPa)
Platinum 10: 9 psi ± 0.5 psi (62.0 kPa ± 3.45 kPa) |
| Flow Range: | Platinum 5 and XL - 0.5 to 5 L/min For flowrates less than 1 L/min, we recommend the use of the Invacare Pediatric Flowmeter IRCPF16 Accessory.
Platinum 10 - 2 L/min to 10 L/min |
| Low Flow Alarm: | Platinum 5 - 0 L/min to 0.5 L/min, Platinum 10 - 0 L/min to 1 L/min Rapid Audible Alarm Beeping (No Accessories Connected) |
| Average Power Consumption: | Platinum 5 and XL: 400W Platinum 10: 585W |
| Pressure Relief Mechanism Operational at: | 35 psi ± 3.5 psi (241 kPa ± 24.1 kPa) |
| Change in maximum recommended flow when back pressure of 7kPa is applied: | 0.7 L/min |
| Filters: | Cabinet (2), Outlet HEPA and Compressor Inlet. |
| Safety System: | Current overload or line surge shutdown. High temperature compressor shutdown. High Pressure Alarm w/compressor shutdown. Low Pressure Alarm w/compressor shutdown. Battery Free Power Loss Alarm. SensO2 Oxygen System (SensO2 Model) Low Flow Alarm. |
| Width: | 18 3/8 in ± 3/8 in (46.7 cm ± 1 cm) |
| Height: | 26 3/8 in ± 3/8 in (67.0 cm ± 1 cm) |
| Depth: | 14 3/8 in ± 3/8 in (36.5 cm ± 1 cm) |
| Weight: | Platinum 5 and XL: 52 lbs ± 2 lbs (23.6 kg ± 1 kg)
Platinum 10: 53 lbs ± 2 lbs (24.0 kg ± 1 kg) |
| Shipping Weight: | Platinum 5 and XL: 58 lbs ± 2 lbs (26.3 kg ± 1 kg)
Platinum 10: 60 lbs ± 2 lbs (27.2 kg ± 1 kg) |
| Operating Ambient Temperature: | 50°F - 95°F (10°C - 35°C) at 20-60% relative humidity |
| Exhaust: | Platinum 5 and XL: Less than Ambient + 35°F (+19°C).
Platinum 10: Less than Ambient + 45°F (+25°C) |
| Oxygen Output Temperature: | Platinum 5 and XL: Less than Ambient + 4°F (+2°C)
Platinum 10: Less than Ambient + 8°F (+4.4°C) |
| Cabinet: | Impact Resistant flame-retardant plastic cabinet that conforms to UL 94-V0 |
| Regulatory Listing: | Double Insulated Product ETL certified complying with UL 1097 ETL certified to CSA C22.2 No. 68 |
| Electrical: | No extension cords. |
| Placement: | No closer than three inches from any wall, furniture, draperies, or similar surfaces. |
| Tubing: | 7 ft cannula with a maximum 50 ft of Crush-Proof Tubing (DO NOT pinch).
Platinum 10: Recommended use up to 50 ft high flow tubing with high flow cannula at all flow rates. |
| Relative Humidity: | 20 to 60% |
| Time of Operation: | Up to 24 hours per day. |
| Recommended Storage and Shipping Temperature: | -20°F to 150°F (-29°C to 65°C), at 15-95% relative humidity |