| POWER INPUT: | 24 VDC @ 7.5 Amp or 11-16 VDC @ 10.0 Amp max (12.6 VDC Nom.±td> |
EXTERNAL POWER SUPPLY INPUT: AC POWER SUPPLY:
DC POWER SUPPLY: | 120 VAC, 50/60 Hz @ 2.5 amps 230 VAC, 50/60 Hz @ 1.25 amps
11-16 VDC, (12.6 VDC Nom. @ 10.0 Amp max±td> |
| SOUND LEVEL: | < 40 dBA weighted @ 2 LPM continuous and all pulse settings (1-5±td> |
| ALTITUDE: | Up to 10,000 ft (3046 m) above sea level Titration recommended for use above 10,000 ft (3046 m±td> |
OXYGEN CONCENTRATION:* *BASED ON AN ATMOSPHERIC PRESSURE OF 14.7 PSI (101 KPA) AT 70°F (21°C) NOMINAL | 87% to 95.6%, after initial warm-up period (approximately 5 minutes) at all flow rates |
CONSERVER TRIGGER SENSITIVITY: FACTORY SET - NO ADJUSTMENT PRESSURE ACTIVATED | < 0.20 cmH2O max pressure drop
All settings - using 7 ft (2.1 m) cannula |
CONSERVER TRIGGER DELAY: FACTORY SET - NO ADJUSTMENT | 35 mSec max
Nominal value - using 7 ft (2.1 m) cannula |
CONSERVER BOLUS DELIVERY: FIXED MINUTE VOLUME | Setting 1: 400 cc Setting 2: 800 cc Setting 3: 1200 cc Setting 4: 1600 cc Setting 5: 2000 cc
Total cc's ± 75 cc delivered per minute |
| CONSERVER BREATH RATE CAPACITY: | Up to, and including, 35 BPM without reduction of bolus minute volume |
| CONTINUOUS FLOW SETTINGS: | 0.5 through 3.0 LPM @ 0 psi 0.5 LPM increments All settings are ± 0.2 LPM (2.0 LPM max w/Ext DC Power Cable±td> |
| MAXIMUM RECOMMENDED FLOWRATE WITH 7KPA (1.01 PSI) BACKPRESSURE: | 3.0 LPM |
| PRESSURE RELIEF ACTIVATION: | 20 psi ± 5 psi (137.8 kPa ± 34.5 kPa±td> |
| MAX OUTLET PRESSURE @ 3.0 LPM CONTINUOUS: | 12.0 psig |
DIMENSIONS: (WITHOUT CART) | 16.5 in high x 11 in wide x 8 in deep (41.9 cm high x 27.9 cm wide x 20.3 cm deep) |
WEIGHT: (UNIT WITHOUT CART) | < 20 lbs (9.09 kg) |
BATTERY DURATION: (TIMES ARE APPROXIMATE) |
| Pulse Mode | Continuous Mode |
| Setting 1 = 4.5 hrs | Setting 0.5 LPM = 4.5 hrs |
| Setting 2 = 3.5 hrs | Setting 1 LPM = 3.5 hrs |
| Setting 3 = 3.0 hrs | Setting 2 LPM = 2.5 hrs |
| Setting 4 = 2.5 hrs | Setting 3 LPM = 1.5 hr |
| Setting 5 = 2.5 hrs | |
|
| BATTERY RECHARGE TIME: | 5 hours Recharge time increases if battery is charging while unit is running. |
HUMIDITY RANGE: OPERATING HUMIDITY: TRANSPORT AND STORAGE | 15% to 60% non condensing Up to 95% non condensing |
STANDARD TEMPERATURE RANGE (ALL POWER SOURCES)
OPERATING TEMPERATURE: TRANSPORT AND STORAGE TEMPERATURE: | 41°F to 95°F (5°C to 35°C) -2°F to 140°F (-20°C to 60°C) |
EXTENDED TEMPERATURE RANGE: (USING AC OR DC POWER) OPERATING TEMPERATURE: AC SUPPLY: DC SUPPLY: | 95°F to 104° (35°C to 40°C) Unlimited use all settings, all modes Unlimited use all settings with pulse mode Limited to 2.0 LPM, or less, with continuous mode |
EXTENDED TEMPERATURE RANGE: (USING BATTERY) OPERATING TEMPERATURE: |
95°F to 104° (35°C to 40°C)
| Pulse Mode | Continuous Mode |
| Settings 1, 2 and 3 = unlimited use | 0.5 thru 1.5 LPM = unlimited use |
| Setting 4 = 45 minutes | 2.0 thru 2.5 LPM = 45 minutes |
| Setting 5 = 30 minutes | 3.0 LPM = 30 minutes |
|
| Recommendations for Optimal Performance |
| Operating Temperature: | 41°F to 95°F (5°C to 35°C) |
| Relative Humidity: | 15% to 60% |
| Transport/Storage Temperature: | -2°F to 140°F (-20°C to 60°C) Allow unit to warm, or cool, to operating temperature range before using. |
| Electrical: | No extension cords. |
| Altitude: | Up to 10,000 ft (3046 m) above sea level. |
| Tubing and Cannula: | 7 ft (2.1 m) crush resistant cannula (DO NOT pinch). |
| Environment: | Smoke, pollutant and fume free. No confined spaces (example: no closets). |
| Time of Operation: | Up to 24 hours per day when connected to AC or DC power. |